Cryogen-free dilution refrigerator equipped with DC and RF wiring, circulators, and HEMT RF amplifiers
Main characteristics:• Base temperature: approximately 15 mK (with all presently installed wiring)• Mixing chamber cooling power at 100 mK: 250 uW• Cooling power at 4.2K: >0.5 W• Cooldown time to base temperature: more than 24 hoursMeasurement Circuitry:
• 2 independent sample holders for rf transmission measurements in an approximately 100 MHz bandwidth centered at 600 MHz
• 2 coaxial high frequency output lines with SMA connectors, with superconducting Nb or NbTi cables, circulators, and HEMT amplifiers at the 4 K flange. Measurement bandwidth approximately 100 MHz centered at 600 MHz• up to 6 coaxial input lines up to 18 GHz as well as DC-bias lines
• each sample holder equipped with In-shielding against stray microwaves, and up to 12 Thermocoax-filtered DC lines • 600 MHz Josephson Parametric Amplifier at the mixing chamber flange
• Primary Coulomb Blockade Thermometer (CBT) thermalized to the mixing chamber flange