The TOC instrument measures the amount of total carbon, inorganic carbon and total organic carbon in water.
Installation: 2010
Analyte: TC, IC and TOC in water (standard SFS-ISO 8245)
Measurement principal: 680 oC analytically-aided combustion oxidation/non-dispersive infrared detection (NDIR)
Measuring range:
Detection limit
Acid addition and sparging possible for NPOC analysis
Sample injection volume 10 – 2 000 µL
Sample dilution with in syringe, dilution factor 2 – 50 times
Vial rack : 93 vials (24 mL vial)